DDA 5005A, DDA 3000 DDA 5005A XXL
Quick Specs
- - 3 GHz, 5 GHz Bandwidth
- - 10 GS/s Standard Sample Rate
- - 24 Mpts Memory Standard
 |
|
< Back to Disk Drive Analyzer Home
Specifications
Vertical
System
|
DDA
3000
|
Analog
Bandwidth
@ 50Ω (-3 dB) |
3
GHz |
| Rise Time (Typical) |
150 ps |
| Input
Channels |
4 |
| Bandwidth
Limiters |
25 MHz; 200 MHz; 1 GHz |
| Input
Impedance |
50Ω or 1 MΩ || 15 pF; 10 MΩ || 11 pF with PP005A Probe |
| Input
Coupling |
1 MΩ: AC, DC, GND; 50 Ω: DC |
| Maximum
Input Voltage |
50 Ω: 5
Vrms,
1 MΩ: 100 Vmax (peak AC: ≤ 5 KHz + DC) |
| Channel-Channel Isolation |
250:1 at same V/div setting, 40:1 at 3 GHz |
| Vertical
Resolution |
8
bits; up to 11 bits with
enhanced resolution (ERES) |
| Sensitivity |
50 Ω: 2 mV – 1 V/div fully variable; 1 MΩ: 2 mV – 2 V/div fully variable |
| DC Gain Accuracy |
±1.5% of full scale; ±1% (typical) |
Offset
Range |
50 Ω: ±700 mV @ 2–4.95 mV/div
±1.5 V @ 5–100 mV/div
±10 V @ .102–1 V/div
1 MΩ: ±700 mV @ 2–4.95 mV/div
±1.5 V @ 5–100 mV/div
±20 V @ 0.102–2 V/div |
| Offset Accuracy |
±(1.5% of full scale + 0.5% of offset value + 2 mV) |
| Horizontal
System |
| Timebases |
Internal
timebase common to 4 input channels; An external clock
may be applied at the Auxiliary input |
| Time/Division Range |
20 ps/div – 10 s/div |
| Clock
Accuracy |
≤ 5
ppm |
| Time Interval Accuracy |
≤ 0.06 / SR + (5 ppm * Reading)
(rms) |
| Sample Rate & Delay Time Accuracy |
±10 ppm ≤ 10 s interval |
| Jitter Noise Floor |
2 ps rms @ 100 mV/div (typical) |
| Trigger and Interpolator Jitter |
≤ 2.5 ps (typical) |
| Channel-Channel Deskew Range |
±9 x time/div setting, 100 ms max, each channel. |
| External
Clock |
30 MHz – 1 GHz; 50 Ω impedance; applied at the auxiliary input |
| Acquisition
System |
| Single-Shot
Sample Rate/Ch |
10
GS/s |
| 2
Channel Max |
20
GS/s |
| Random
Interleaved Sampling (RIS) |
200
GS/s for repetitive signals:
20 ps/div – 1 µs/div |
| Maximum
Trigger Rate |
150,000
waveforms/second (in Sequence Mode, up to 4 channels) |
| Intersegment
Time |
6 µs |
| Maximum
Acquisition Points/Ch |
50 Mpts, 100 Mpts in two channel mode. |
| Maximum nummber of segments in Sequency Mode: |
25,000 segments |
| |
4
Ch / (2 Ch) |
Sequence
Mode |
| Standard |
48M
/ 100M |
25,000
segments |
|
| Acquisition
Processing |
| Averaging |
Summed or Continuous averaging up to 1 million sweeps |
| Enhanced
Resolution (ERES) |
From 8.5
to 11 bits vertical resolution |
| Envelope
(Extrema) |
Envelope,
floor, or roof for up to 1 million sweeps |
| Interpolation |
Linear or Sin x/x |
| Triggering
System |
| Modes |
Normal,
Auto, Single, and Stop |
| Sources |
Any
input channel, External, Ext X10, Ext/10, or line; slope
and level unique to each source (except line trigger) |
| Coupling |
DC |
| Pre-trigger
Delay |
0-10% of horizontal time scale |
| Post-trigger
Delay |
0 –
10,000 divisions |
| Hold-off
by Time or Events |
up to 20 s or from 1 to 99,999,999 events |
| Internal
Trigger Range |
±5 div
from center |
| Max
Trigger Frequency |
Trigger Senstivity (Edge) 2 div < 3 GHz, 1 div < 2 GHz |
|
Max. Trigger Frequency, SMART Trigger | 750 MHz ≥ 10 mV |
| Automatic
Setup |
| Auto
Setup |
Automatically
sets timebase, trigger, and sensitivity to display a
wide range of repetitive signals. |
| Vertical
Find Scale |
Automatically
sets the vertical sensitivity and offset for the
selected channels to display a waveform with maximum
dynamic range. |
| Probes |
| Probes |
(2) PP005A standard; Optional passive and active probes available. |
| Probe
System: Probus |
Automatically
detects and supports a variety of compatible probes. |
| Scale
Factors |
Automatically
or manually selected depending on probe used. |
| Color
Waveform Display |
| Type |
Color
10.4" flat-panel TFT-LCD with high resolution touch
screen |
| Resolution |
SVGA;
800 x 600 pixels |
| Number
of Traces |
Display
a maximum of 8 traces. Simultaneously display channel,
zoom, memory, and math traces. |
| Grid
Styles |
Auto, Single,Dual, Quad,Octal, XY, Single + XY,Dual + XY |
| Waveform
Styles |
Sample
dots joined or dots only |
| Analog
Persistence Display |
| Analog
& Color-Graded Persistence |
Variable
saturation levels; stores each trace’s persistence
data in memory |
| Persistence
Selections |
Select
analog or color, or three-dimensional. |
| Trace
Selection |
Activate Persistence on all or any combination of traces |
| Persistence
Aging Time |
Select
from 500 ms to infinity |
| Sweeps
Displayed |
All
accumulated, or all accumulated with last trace
highlighted |
| Zoom
Expansion Traces |
| Display up to 8 Math/Zoom traces |
| CPU |
| Processor |
Intel Pentium® P4, 2.54 GHz or better, 2 GB memory |
| Operating System |
Microsoft Windows® XP Professional |
| Real-time
Clock |
Dates,
hours, minutes, seconds displayed with waveform. SNTP support to synchronize to precision internet clocks. |
| Internal
Waveform Memory |
| M1,
M2, M3, M4 Internal Waveform Memory (Store full-length
waveforms with 16 bits/data point) or store to any
number of files limited only by data storage media. |
| Setup
Storage |
| Front
Panel and Instrument Status |
Store
to the internal hard drive, or to a USB-connected peripheral device. |
| Interface |
| Remote
Control |
Via Windows Automation, or via LeCroy Remote Command Set |
| GPIB
Port |
Supports
IEEE – 488.2 |
| Ethernet
Port |
10/100BaseT
Ethernet interface |
| USB
Ports |
6 USB 2.0 ports, (2 accessable on front panel) support Windows® compatible devices |
| External
Monitor Port Standard |
15-pin
D-Type SVGA-compatible |
| Parallel
Port |
1
standard |
| Auxiliary
Output |
| Signal
Types |
Select
from calibrator or control signals output on front panel |
| Coupling |
50 Ω: DC; 1 MΩ: AC, DC, GND |
| Max. Input Voltage |
50 Ω: 5 V rms; 1 MΩ: 250 V (Peak AC < 10 kHz + DC) |
| Calibrator
Signal |
5 Hz – 1 MHz square wave or DC level;
0.0 to 5.0 V into 50 Ω (0-1 V into 1 MΩ)
or TTL volts (selectable) |
| Control
Signals |
Trigger
ready, trigger out, pass/fail status |
| Auxiliary
Input |
| Signal
Types |
Select
from External Trigger or External Clock input on front
panel |
| General |
| Auto
Calibration |
Ensures
specified DC and timing accuracy is maintained for 1
year minimum |
| Max. power consumption |
650 W/ 650 VA
200–240 V AC at 50/60 Hz;
Automatic AC Voltage selection
Power
consumption: < 800VA |
| Environmental |
| Temperature (Operating) |
+5 °C to +40 °C including CD-ROM drive |
| Temperature (Non-Operating) |
-20°C to +60°C |
| Humidity (Operating) |
5% to 80% relative humidity (non-condensing) up to +30 °C.
Upper limit derates to 25% relative humidity (non-condensing) at +40°C |
| Humidity (Non-Operating) |
5% to 95% relative humidity (non-condensing) as tested per MIL-PRF-28800F |
| Altitude (Operating) |
up to 10,000 ft (3048 m) at or below +25°C |
| Altitude (Non-Operating) |
up to 40,000 ft (12,192 m) |
| Random Vibration (Operating) |
0.31 g rms 5 Hz to 500 Hz, 15 minutes in each of three orthogonal axes |
| Random Vibration (Non-Operating) |
2.4 g rms 5 Hz to 500 Hz, 15 minutes in each of three orthogonal axes |
| Functional Shock |
20 g peak, half sine, 11 ms pulse, 3 shocks (positive and negative) in each of three orthogonal axes, 18 shocks total |
| Physical
Dimensions |
| Dimensions
(HWD) |
264 mm x
397 mm x 491 mm; 10.4" x 15.6" x 19.3"
(height excludes feet) |
| Weight |
18 kg;
39 lbs |
| Shipping
Weight |
24 kg;
53 lbs |
| Certifications |
| CE compliant, UL and cUL listed; conforms to EN 61326-1, EN 61010-1, UL 3111-1, and CSA C22.2 No. 1010.1 |
| Warranty
and Service |
| 3-year
warranty; calibration recommended annually |
| Optional
service programs include extended warranty, upgrades,
and calibration services. |
| Basic
Triggers |
| Edge/Slope/Line |
Triggers
when signal meets slope and level condition. |
| SMART®Triggers |
| State
or Edge Qualified |
Triggers
on any input source only if a defined state or edge
occurred on another input source. Delay between sources
is selectable by time or events. |
| Dropout |
Triggers
if signal drops out for longer than selected time
between 2 ns and 20 s. |
| Pattern |
Logic
combination (AND, NAND, OR, NOR) of 5 inputs
(4 channels
and external trigger input)
Each source can be high, low, or don’t care.
Trigger
at start or end of the pattern. |
| SMART
Triggers with Exclusion Technology |
| Glitch |
Triggers
on positive or negative glitches with widths selectable
from 600 ps to 20 s or on intermittent faults. |
| Signal
or Pattern Width |
Triggers
on positive or negative pulse widths selectable from 600
ps to 20 s or on intermittent faults. |
| Signal
or Pattern Interval |
Triggers
on intervals selectable between 2 ns and 20 s. |
| Disk
Drive Triggers |
| Sector |
Trigger
on the n'th sector pulse after index. Index and sector
pulse polarity and sector pulse number are selectable. |
| Servo
Gate |
Trigger
on the n'th servo gate after index and every m'th
thereafter. Index and servo gate pulse polarity are
selectable. |
| PES
Trigger |
Trigger
on Position Error Signal (PES) exceeding an adjustable
voltage window. Servo gate can be selected as qualifier. |
| Read
Gate Trigger |
Trigger
on any read gate longer than an adjustable Sector ID
filed length. |
Math
Tools
Display up to 8 math function traces (F1 – F8). The
easy-to-use graphical interface simplifies setup of up to two
operations on each function trace. Function traces can be
chained together to perform math-on-math.
| absolute
value |
log
(base 10) |
| average
(summed) |
negate |
| average
(continuous) |
product
(x) |
| difference
(–) |
ratio
(/) |
| differentiate |
reciprocal
(invert) |
enhanced
resolution
(to 11 bits vertical) |
resample
(deskew) |
| envelope |
rescale
(with units) |
| exp
(base e) |
roof |
| exp
(base 10) |
(sin)
x/x |
| FFT |
square |
| floor |
square
root |
| identity |
sum (+) |
| integrate |
trend (datalog) |
| log
(base e) |
Auto-correlation |
|
Histogram |
FFT
includes: power averaging, power density, real and imaginary
components, and frequency domain parameters.
Pass/Fail
Testing
Test waveforms by comparing their shape to test templates, and simultaneously check multiple parameters versus selectable parameter or mask limits.Pass or fail conditions can initiate actions including document to local or networked files, or e-mail the image of the failure,save waveforms, or send a GPIB SRQ, or pulse to trigger another device.
Automated
Disk Drive Measurements
| TAA |
lbase |
ltmn |
msnr |
| TAA+ |
lbsep |
ltmx |
rsnr |
| TAA- |
lmax |
ltot |
m_to_r |
| PW50 |
lmin |
ltpt |
nbph |
| PW50+ |
lnum |
lttp |
nbpw |
| PW50- |
lpp |
ltut |
|
| Resolution |
ltbe |
NLTS |
|
| Overwrite |
ltbp |
ACSN |
|
Standard
Automated Measurements
| amplitude |
peak-to-peak |
| area |
period |
| base |
risetime |
| cycles |
rms |
| cycle
std. deviation |
std.
deviation |
| cycle
mean |
top |
| cycle
median |
width |
| cycle
rms |
last |
| data |
median |
| delay |
number
of points |
| duty
cycle |
phase |
| duration |
time
@ minimum (min.) |
| falltime |
time
@ maximum (max.) |
| frequency |
Δ
delay |
| first |
Δ
time @ level |
| maximum |
Δ
time @ level from trigger |
| mean |
Δ
time from clock to data + (setup time) |
| minimum |
Δ
time from clock to data- (Hold time) |
| +overshoot |
18
Histogram Parameters |
| –overshoot |
|
Jitter
measurement for parameters including: period, cycle-cycle,
frequency, and edge@lv, with JitterTrack up to 200 edges.
Advance
Drive Analysis
Advance Drive Analysis capabilities of the DDA3000 include:
Additional
waveshape analysis capabilities include:
-
FFT
capability includes: power averaging, power density, real
and imaginary components and frequency domain parameters
-
Parameter
Math - add, subtract, multiply or divide two different
parameters
-
User-definable
parameter measurements
-
User-definable
math functions
|